OptiStar® Elite technical specifications

Simple, battery-free
Use with prefilled syringe and vials

Installation
Pedestal; vehicle suspended arm (US only)

Flow rate
A Side: 60 mL syringes: 0.1–10.0 mL/sec; 10, 15, 20, 30 mL syringes: 0.1–8.0 mL/sec.
B Side: 60 mL single-use disposable syringes: 0.1–8.0 mL/sec; 50, 125 mL prefilled saline syringes*: 0.1–8.0 mL/sec

Pressure Limit
60 mL empty syringes: 150 psi; 10, 15, 20, 30 mL prefilled syringes: 200 psi*; 50, 125 mL prefilled saline syringes: 150 psi*

Syringe Sizes
60 mL empty syringes; 10, 15, 20, 30 mL prefilled syringes*; 50, 125 mL prefilled saline syringes*

Volume
A Side: 10, 15, 20 mL syringes: 1 mL to 20 mL; 30, 60 mL syringes: 1 mL to installed syringe size.
B Side: 60 mL syringes: 1 mL to installed syringe size; 50, 125 mL prefilled saline syringes*: 1 mL to installed syringe size*

Scan Delay    0-60 seconds

Inject Delay    0-300 seconds

Number of Phases    6 including Timing Bolus® feature

Phase Delay    0-600 seconds + pause

Number of Programs    40

Inject History    Last 48 injected protocols

MR Compatibility    Up to 3 Tesla

*where available

Featuring:

  • Patency Check®: test vein prior to injection (dedicated powerhead key)
  • Timing Bolus®: test injection to confirm protocol
  • Variable drip mode: ensure IV line patency throughout the procedure
  • Pressure automatic sensing & setting to insure best security
  • Protocol: up to 6 phases with Timing Bolus®, even more with injection & phase delays
  • ‘One click loading’ / Auto retract rams for quick turnaround
  • Console Enable
  • Powerhead keys
  • Colour touchscreen, uncluttered display
  • Clear indicators & status line